Gauge pressure transmitter - high pressure - 70 MPa - FKR

High gauge pressure measurement

Fields of application

Designed for pressure monitoring in critical industrial applications such as oil & gas, cryogenics, hydraulic presses and test benches.

Features

  • High accuracy: ±0.065% of set measurement range
  • Measures very high pressures up to 70 MPa (700 bar / 10,152 psi)
  • 4-20 mA output with HART™ communication 7
  • Robust materials (Inconel® 625, Inox 318LN and Inox 316L) and design for extreme environments
  • SIL2 / SIL3 certification and ATEX/IECEx compatibility
  • No-open configuration thanks to magnetic buttons
Reference: FKR-06 Please contact us

Specifications

The FKR gauge pressure transmitter is specially designed for industrial applications requiring reliable and accurate measurement of very high pressures, up to 70 MPa, i.e. 700 bar, 70,000 kPa or approx. 10,152 psi. Thanks to its piezoresistive silicon transducer and digital processing electronics, it offers excellent long-term stability and outstanding accuracy. It meets the most stringent safety requirements (SIL2/SIL3) and integrates easily into control systems thanks to its 4-20 mA output with HART™ protocol. Robust and adaptable, it is perfectly suited to harsh environments and critical processes under high stress, notably in hydrogen production systems, industrial hydraulic presses, plastic injection machines, supercritical steam plants, hydrotreating or hydrocracking units, hydraulic control of subsea valves, high-pressure cryogenics, high-pressure systems in civil nuclear power and CO₂ storage and distribution.

 

TypeIntelligent 4-20 mA Pressure transmitter with HART™ 7 protocol
HousingL- or T-shaped housing
Measuring ranges [e.max.]From 4.375 to 70 MPa
Measurement limit-100 to 70,000 (kPa)
Permissible overpressure105 MPa
Weight per kg (without indicator)1,5
AccuracyFrom 1/10 to 1/1 of URL: ±0.065 % of span
From 1/16 to 1/10 of URL: ±(0.015 + 0.05 × 0.1 × URL/span) % of span
Materials in contact with the fluidMembraneInconel® 625 (with or without gold plating)
ConnectionStainless steel 318LN or Inconel® 625
Sensor bodyStainless steel 316L or Inconel® 625
Non-contact materialsHousingCast aluminum (polyester coating) or 316L stainless steel
Filling fluidSilicone or mineral oil
Stability±0.1% of upper range limit for 5 years
Elevation / DeletionZero offset: ±0.2% of max. scale (URL)
Scale setting rangeFrom 1 to 1/16 of maximum scale
Measurement cycle40 ms
Communication protocolHART™ 7 communication protocol only
Response timeTime constant: 80 ms
Dead time: approx. 60 ms
Total time ≈ 140 ms
Process temperatureFrom -40 to 100°C
Ambient temperatureFrom -40 to +85°C
-20 to +80°C (with optional LCD unit)
-40 to +60°C (with optional lightning arrester)
Storage temperatureFrom -40 to +90°C
Temperature effectsZero offset: ±(0.075 + 0.125 × URL/extent) % / 28 °C
Total effect: ±(0.095 + 0.0125 × URL/extent) % / 28 °C
Power supply10.5 to 45 V DC
Output signal / Load resistance4-20 mA / 250 Ω typical + HART™ (NAMUR NE 43 compliant)
Dielectric strength500 V AC for 1 min (between circuit and earth)
Insulation resistance> 100 MΩ at 500 V DC (between circuit and ground)
DampingConfigurable from 0.04 to 32 seconds
Zero and span adjustmentVia external sensor screw, local three-button indicator or HART™ protocol
Electrical connectionsM20×1.5" / 1/2 -14 NPT / Pg13.5
AssemblyDirect or Ø60.5 mm tube mounting (U-bolt stainless steel 304L/316L)
Process connectionAutoClave F250C (F) or F250C (M) (adapter) or 1/4-18 NPT (F) (adapter)
CompatibilityCEM : EN 61326-1, EN 61326-2-3, EN 61326-3-1
ATEX/ IECEx / cCSAus : Intrinsically safe (Ex ia), Explosion-proof enclosure (Ex d), XP/IS/NI Division 1 and 2
OptionsDigital indicator, degreasing treatment, oxygen service, chlorine service, additional stainless steel identification plate

Documentations

Datasheet: High gauge pressure transmitter FCX-AIV V6 FKR

Datasheet Download

Brochure : Pressure transmitters

Brochures Download

Brochure: Measures for energy and the environment

Brochures Download

Gauge pressure transmitter - high pressure - 70 MPa - FKR