[ Printer-friendly version ][ Send to friend ]The non dispersive infrared ZRJ, ZKJ, ZRE and the portable ZSV are intended for the continuous measurement of the concentrations of carbon monoxide and carbon dioxide (CO and CO2), sulfur dioxide (SO2), nitrogen oxides (NOx), methane (CH4) and oxygen (oxide of zirconium, electrochemical or paramagnétique) in the gas emissions of the fixed sources of pollution (boilers, ovens...).
The non dispersive infrared ZRJ, ZKJ, ZRE and the portable ZSV are intended for the continuous measurement of the concentrations of carbon monoxide and carbon dioxide (CO and CO2), sulfur dioxide (SO2), nitrogen oxides (NOx), methane (CH4) and oxygen (oxide of zirconium, electrochemical or paramagnétique) in the gas emissions of the fixed sources of pollution (boilers, ovens...).
They allow the measurement from 1 to 5 gases. The measurement system in simple or double beam and unique source gets to the user a maintenance of the easiest and assures a very good precision and a very big stability of the measure. The wide back-lit screen allows the display of the measures and the access the various parameters for configuration and maintenance. They possess very numerous functions: automatic calibration for zero and span, alarms, automatic change of range, auto-diagnoses, correction by the dry oxygen, a digital connection with the protocol Modbus.
ZSV portable type

ZRE model | |
| ![]() |
| Measurement principle | NO, SO2, CO, CO2, CH4: non dispersive infrared ray absorption (single beam system) O2: fuel cell (built in) or zirconia (externally installed ZFK7 by Fuji) |
| Measurable component and range | Measeured gas Mini range Maxi range |
| Stability | ±0.5% FS |
| Linearity | ±1.0% FS |
| Zero drift | Within ±2% FS /week |
| Span drift | Within ±2% FS /week |
| Response time | Within 60 sec (90% response from gas inlet). Varies depending on the components to be measured and the measurement range. |
| Analog output signal | 4 to 20 mA or 0 to 1 VDC (12 points at max.) Instantaneous value output (Concentration of each gas component measured) Option: O2 correction instantaneous value output, O2 correction average output, O2 average output |
| Display | LCD with backlight (Japanese or English by designation) |
| Range switching | Manual switching by key operation, auto switching, external contact input switching (option) |
| External digital input (option) | Voltage contact (supply 12 to 24 VDC/15 mA max. at ON) 9 points at max. Range switching, auto calibration start, output signal hold, average value reset |
| Contact output function (option) | 1c relay contact (15 points at max.) Identification of each component range, instrument failure,calibration error, auto calibration in progress, upper / lower limit alarm for each component, pump ON/OFF, solenoid valve drive for auto calibration |
| Communication function (option) | RS-485 (MODBUS protocol) Details of communication: Read/write of each setting, output of measured concentration and instrument status, Type-B with USB connector (front face) and USB driver |
| Sample gas flowmeter | Built in |
| Gas outlet/inlet dimension | Rc1/4 or NPT1/4 |
| Purge gas flow rate | 1 L/min (Performed as required.) |
| Structure | Indoor type with steel case |
| Ambient temperature | −5°C to 45°C |
| Humidity | 90 RH or lower (No condensation allowed) |
| Mounting method | 19” rack mount, panel mount, desktop |
| Power supply voltage | 100 to 240 VAC, 50/60 Hz, 100VA |
| Outside dimensions | 133×483×418 mm (19” rack mount) 133×440×418 mm (Panbel mount) |
| Mass | Approx. 8 kg (5 - components analyzer) |
| Applicable standard | CE mark |
| Flow Rate | 0.5 L/min ±0.2 L/min |
| Temperature | 0°C to 50°C |
| Pressure | 10 kPa maxi |
| Dust | 100 μg/Nm3 or lower (Particle size: 0.3 μm or smaller) |
| Mist | Not allowed |
| Moisture | Saturation at room temperature or lower (No condensation allowed.) Saturation at 2°C or lower (No condensation allowed.) |
| Corrosive component | HCl : 1 ppm maxi |
ZKJ model : Ideal for measurement of low-concentration components | |
Simultaneous ansd continuous measurement of gas concentration of up 5 components out of CO2, CO, CH4, NOx, SO2 and O2
| ![]() |
| Measurement principle | NOx, SO2, CO, CO2, CH4 : Non-dispersive infrared ray system (Double-beam) |
| Measured component | Measured gas Mini range Maxi range NOx: 0 to 50ppm 5000ppm SO2: 0 to 50ppm 10% CO: 0 to 50ppm 100% CO2: 0 to 20ppm 100% CH4: 0 to 200ppm 100% O2: 0 to 5% 25% (2-range switching, Maximum range ratio 1:5, O2 excluded) |
| Stability | ±0.5%FS (±1%FS for concentration of less than 50ppm) |
| Linearity | ±1.0% FS or lower |
| Zero drift | ±1.0%FS or lower/week (±2.0%FS / week for concentration from 50ppm to 200ppm) |
| Span drift | ±2.0%FS or lower/week (±2.0%FS / day for concentration of less than 50ppm) |
| Response time | 90% response from gas inlet: 60 sec. or shorter |
| Output signal | 4 to 20mA DC or 0 to 1V DC (Max. non-insulated output point: 12) Instantaneous output value (measured gas concentration of each component) Instantaneous output value after O2 correction, Average output value after O2 correction, Average O2 output Permissible load resistance: 550Ω or lower (4 to 20mA DC) 100kΩ (0 to 1V DC) |
| External contact input | No voltage contact |
| Sorties contact | Identification échelle pour chaque gaz, Défaut analyseur, Défaut étalonnage, Etat d’étalonnage auto, Alarme compteur dépassement, Valeur instantanée pour chaque gaz après correction, Marche/Arrêt pompe, Arrêt alimentation |
| Communication function | RS-232C (MODBUS) option |
| Auto calibration function | Auto zero and span calibration (Calibration cycle settable) |
| Display | LCD with backlight |
| Power supply voltage | 100 to 240 VAC, 50/60 Hz, 250VA |
| Outside dimensions | 177x 483 x 578 mm |
| Weight | Approx. 22kg (5 gaz) |
| Temperature | 0°C to 50°C |
| Pressure | 10kPa or lower (The gas outlet should be at atmospheric pressure) |
| Dust | 100μg/Nm3 or lower with particle size of 1μm or lower |
| Mist | No mist allowed |
| Moisture | Saturated at 2°C (No condensation allowed) |
| Corrosive component | 1ppm or lower |
ZRJ model : Ideal for combustion control of various industrial furnaces | |
Simultaneous and continuous measurement of gas concentration : 4 gas of CO2, CO, CH4, NOx, SO2 and O2
| ![]() |
| Measurement principle | NOx, SO2, CO, CO2, CH4: Non-dispersive infrared ray system (single-beam) |
| Measured component | Measured gas Mini range Maxi range NOx: 0 to 500ppm 5000ppm SO2: 0 to 500ppm 5000ppm CO: 0 to 200ppm 100% CO2: 0 to 500ppm 100% CH4: 0 to 1000ppm 100% O2: 0 to 5% 25% (2-range switching, Maximum range ratio 1:5, O2 excluded) |
| Stability | ±0.5% FS |
| Linearity | ±0.1% FS or lower |
| Zero drift | ±2.0%FS or lower/week |
| Span drift | ±2.0%FS or lower/week |
| Response time | 90% response from gas inlet: 15 sec. or shorter (2-component measurement) |
| Output signal | 4 to 20mA DC or 0 to 1V DC (Max. non-insulated output point : 8) Instantaneous output value (measured gas concentration of each component) Instantaneous output value after O2 correction, Average output value after O2 correction, Average O2 output Permissible load resistance: 550Ω or lower (4 to 20mA DC), 100kΩ (0 to 1V DC) |
| External contact input | No voltage contact Auto calibration start, Average value reset, Range selection, Output hold |
| Contact output | Range identification of each component, Instrument error, Calibration error, Auto calibration in progress, CO peak count alarm, Instantaneous value concentration alarm for each component, Pump ON/OFF |
| Communication function | RS-232C (MODBUS) option |
| Auto calibration function | Auto zero and span calibration (Calibration cycle settable) |
| Display | LCD with backlight |
| Power supply voltage | 100 to 240 VAC, 50/60 Hz, 70VA |
| Outline dimensions | 177 x 483 x 493 mm |
| Weight | Approx. 10kg |
| Temperature | 0°C to 50°C |
| Pressure | 10kPa or lower (The gas outlet should be at atmospheric pressure) |
| Dust | 100μg/Nm3 or lower with particle size of 1μm or lower |
| Mist | No mist allowed |
| Moisture | Saturated at 2°C (No condensation allowed.) |
| Corrosive component | 1ppm or lower |