Infrared gas analyzers

Infrared gas analyzers

[ Printer-friendly version ][ Send to friend ]

The non dispersive infrared ZRJ, ZKJ, ZRE and the portable ZSV are intended for the continuous measurement of the concentrations of carbon monoxide and carbon dioxide (CO and CO2), sulfur dioxide (SO2), nitrogen oxides (NOx), methane (CH4) and oxygen (oxide of zirconium, electrochemical or paramagnétique) in the gas emissions of the fixed sources of pollution (boilers, ovens...).
They allow the measurement from 1 to 5 gases. The measurement system in simple or double beam and unique source gets to the user a maintenance of the easiest and assures a very good precision and a very big stability of the measure. The wide back-lit screen allows the display of the measures and the access the various parameters for configuration and maintenance. They possess very numerous functions: automatic calibration for  zero and span, alarms, automatic change of range, auto-diagnoses, correction by the dry oxygen, a digital connection with the protocol Modbus.

ZSV portable type

ZRE model

  • Simultaneous and continuous measurement of 5 components (CO2, CO, CH4, NOx, SO2 and O2).
  • Excellent prolonged stability
  • Virtually unaffected by the interference of moisture
  • Substantial functions, including automatic calibration, communication (RS485) and alarms (option).

 

General specifications

 
Measurement principleNO, SO2, CO, CO2, CH4: non dispersive infrared ray absorption (single beam system)
O2: fuel cell (built in) or zirconia (externally installed ZFK7 by Fuji)
Measurable component and range

Measeured gas     Mini range        Maxi range
NO                         0-200ppm        0-5000ppm
SO2                       0-200ppm        0-10 vol%
CO2                       0-100ppm        0-100 vol%
CO                         0-200ppm        0-100 vol%
CH4                       0-500ppm        0-100 vol%
O2 (built in)            0-10 vol%        0-25 vol%
O2 zirconia            0-5 vol%          0-25 vol%
Switching between 2 ranges allowed for each component. Maxi range ratio: 1:10 (excluding O2)

Stability±0.5% FS
Linearity±1.0% FS
Zero driftWithin ±2% FS /week
Span driftWithin ±2% FS /week
Response timeWithin 60 sec (90% response from gas inlet). Varies depending on the components to be measured and the measurement range.
Analog
output signal
4 to 20 mA or 0 to 1 VDC (12 points at max.)
Instantaneous value output (Concentration of each gas component measured)
Option: O2 correction instantaneous value output, O2 correction average output, O2 average output
Display

LCD with backlight (Japanese or English by designation)
Instantaneous value of each component, O2 correction instantaneous value.
O2 correction average, O2 average, parameter setting, with auto OFF function

Range switchingManual switching by key operation, auto switching, external contact input switching (option)
External digital
input (option)
Voltage contact (supply 12 to 24 VDC/15 mA max. at ON) 9 points at max.
Range switching, auto calibration start, output signal hold, average value reset
Contact output
function (option)
1c relay contact (15 points at max.)
Identification of each component range, instrument failure,calibration error, auto calibration in progress, upper / lower limit alarm for each component, pump ON/OFF, solenoid valve drive for auto calibration
Communication
function (option)
RS-485 (MODBUS protocol)
Details of communication: Read/write of each setting, output of measured concentration and instrument status, Type-B with USB connector (front face) and USB driver
Sample gas
flowmeter
Built in
Gas outlet/inlet
dimension
Rc1/4 or NPT1/4
Purge gas flow rate1 L/min (Performed as required.)
StructureIndoor type with steel case
Ambient
temperature
−5°C to 45°C
Humidity90 RH or lower (No condensation allowed)
Mounting method19” rack mount, panel mount, desktop
Power supply
voltage
100 to 240 VAC, 50/60 Hz, 100VA
Outside dimensions133×483×418 mm (19” rack mount)
133×440×418 mm (Panbel mount)
MassApprox. 8 kg (5 - components analyzer)
Applicable standardCE mark

 

Standard measured gas conditions for gas analyzer

 
Flow Rate0.5 L/min ±0.2 L/min
Temperature0°C to 50°C
Pressure10 kPa maxi
Dust100 μg/Nm3 or lower (Particle size: 0.3 μm or smaller)
MistNot allowed
MoistureSaturation at room temperature or lower
(No condensation allowed.)
Saturation at 2°C or lower (No condensation allowed.)
Corrosive componentHCl : 1 ppm maxi

 

General use model : ZPA, high performance model : ZPB and ultra low concentration measurement : ZPG.

Features
 
1.Wide measurable range from 0 - 5ppm up to 100%
2.Excellent stability: Zero drift ±0.5% FS/week or better(ZPB,ZPG)
3.Simultaneous and continuous measurement of up to 5 gas components(ZPA,ZPB)
4.Compact and lightweight: 130(H)×483(W)×282(D)mm, 11kg or less.
5.Built-in paramagnetic or fuel cell oxygen sensor available
6.Unique simple measurement cell design reduces maintenance
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

General specifications

Minimum measuring range
Measuring gas componentsGeneral use model
(Model:ZPA)
High performance
(Model:ZPB)
Ultra-Low concentration measurement
(Model:ZPG)
NO0 - 200ppm0 - 50ppm0 - 10ppm
SO20 - 200ppm0 - 50ppm0 - 10ppm
CO20 - 100ppm0 - 50ppm0 - 5ppm
CO0 - 200ppm0 - 50ppm0 - 5ppm
CH40 - 500ppm--
O20 - 5%0 - 5%0 - 5%
 ZPAZPBZPG

 

ZKJ model : Ideal for measurement of low-concentration components

Simultaneous ansd continuous measurement of gas concentration of up 5 components out of CO2, CO, CH4, NOx, SO2 and O2

General Specifications

 
Measurement principle

NOx, SO2, CO, CO2, CH4 : Non-dispersive infrared ray system (Double-beam)
O2: Paramagnetic type (built in) or zirconia type (Type ZFK7, Separately installed)

Measured componentMeasured gas     Mini range        Maxi range
NOx:                    0 to 50ppm       5000ppm
SO2:                    0 to 50ppm       10%
CO:                      0 to 50ppm       100%
CO2:                    0 to 20ppm       100%
CH4:                    0 to 200ppm     100%
O2:                      0 to 5%              25%
(2-range switching, Maximum range ratio 1:5, O2 excluded)
Stability±0.5%FS (±1%FS for concentration of less than 50ppm)
Linearity±1.0% FS or lower
Zero drift±1.0%FS or lower/week
(±2.0%FS / week for concentration from 50ppm to 200ppm)
Span drift±2.0%FS or lower/week
(±2.0%FS / day for concentration of less than 50ppm)
Response time90% response from gas inlet: 60 sec. or shorter
Output signal4 to 20mA DC or 0 to 1V DC (Max. non-insulated output point: 12)
 Instantaneous output value (measured gas concentration of each component)
 Instantaneous output value after O2 correction, Average output value after O2 correction, Average
 O2 output
Permissible load resistance:
 550Ω or lower (4 to 20mA DC)
 100kΩ (0 to 1V DC)
External contact
input

No voltage contact
 Auto calibration start, Average value reset, Range selection, Output hold, Pump ON/OFF

Sorties contactIdentification échelle pour chaque gaz, Défaut analyseur, Défaut étalonnage, Etat d’étalonnage auto, Alarme compteur dépassement, Valeur instantanée pour chaque gaz après correction, Marche/Arrêt pompe, Arrêt alimentation
Communication functionRS-232C (MODBUS) option
Auto calibration
function
Auto zero and span calibration (Calibration cycle settable)
Display

LCD with backlight
 Instantaneous value of each component,
 Instantaneous value after O2 correction, Average value after O2 correction, Average O2 value,  CO peak count
Parameter setting display (English or Japanese can be selected)

Power supply voltage100 to 240 VAC, 50/60 Hz, 250VA
Outside dimensions177x 483 x 578 mm
WeightApprox. 22kg (5 gaz)

Standard measured gas conditions for gas analyzer

 
Temperature0°C to 50°C
Pressure10kPa or lower (The gas outlet should be at atmospheric pressure)
Dust100μg/Nm3 or lower with particle size of 1μm or lower
MistNo mist allowed
MoistureSaturated at 2°C (No condensation allowed)
Corrosive component1ppm or lower

 

ZRJ model : Ideal for combustion control of various industrial furnaces

Simultaneous and continuous measurement of gas concentration : 4 gas of CO2, CO, CH4, NOx, SO2 and O2

 

General specifications

 
Measurement principle

NOx, SO2, CO, CO2, CH4: Non-dispersive infrared ray system (single-beam)
O2: Paramagnetic type (built in), galvanic cell type (built in), or zirconia type (Type ZFK7, separately
installed)

Measured
component
Measured gas    Mini range             Maxi range
NOx:                   0 to 500ppm          5000ppm
SO2:                   0 to 500ppm          5000ppm
CO:                     0 to 200ppm         100%
CO2:                   0 to 500ppm         100%
CH4:                   0 to 1000ppm       100%
O2:                     0 to 5%                  25%
(2-range switching, Maximum range ratio 1:5, O2 excluded)
Stability±0.5% FS
Linearity±0.1% FS or lower
Zero drift±2.0%FS or lower/week
Span drift±2.0%FS or lower/week
Response time90% response from gas inlet: 15 sec. or shorter (2-component measurement)
Output signal4 to 20mA DC or 0 to 1V DC (Max. non-insulated output point : 8)
 Instantaneous output value (measured gas concentration of each component)
 Instantaneous output value after O2 correction,
 Average output value after O2 correction, Average O2 output
Permissible load resistance: 550Ω or lower (4 to 20mA DC), 100kΩ (0 to 1V DC)
External contact
input
No voltage contact
Auto calibration start, Average value reset, Range selection, Output hold
Contact outputRange identification of each component, Instrument error, Calibration error, Auto calibration in progress, CO peak count alarm, Instantaneous value concentration alarm for each component, Pump ON/OFF
Communication functionRS-232C (MODBUS) option
Auto calibration
function
Auto zero and span calibration (Calibration cycle settable)
Display

LCD with backlight
 Instantaneous value of each component,
 Instantaneous value after O2 correction, Average value after O2 correction, Average O2 value,    CO peak count
Parameter setting display (English or Japanese can be selected.)

Power supply voltage100 to 240 VAC, 50/60 Hz, 70VA
Outline dimensions177 x 483 x 493 mm
WeightApprox. 10kg

 

Standard measured gas conditions for gas analyzer

 

Standard measured gas conditions for gas analyzer

  
Temperature0°C to 50°C
Pressure10kPa or lower (The gas outlet should be at atmospheric pressure)
Dust100μg/Nm3 or lower with particle size of 1μm or lower
MistNo mist allowed
MoistureSaturated at 2°C (No condensation allowed.)
Corrosive component1ppm or lower

 

[Top of the Page]